Modeling of an implantable device for remote arterial pressure measurement

Abstract: Cardiovascular diseases are the leading causes of illness and death in Europe, having a major impact on healthcare costs. An intelligent stent (e-stent), capable of obtaining and transmitting measurements of physiological parameters, can be a useful tool for real-time monitorization of arterial blockage without patient hospitalization. In this paper, a behavioral model of a pressure sensing-based e-stent is proposed and simulated under several restenosis conditions. Special attention has been given to the need of an accurate fault model, obtained from realistic finite-element simulations, to ensure long-term reliability; particularly for those faults whose behavior cannot be described by usual analytical models.

Otras comunicaciones del congreso o articulos relacionados con autores/as de la Universidad de Cantabria

 Autoría: Miguel J., Lechuga Y., Mozuelos R., Martinez M.,

 Congreso: SPIE Microtechnologies (2013 : Grenoble)

Editorial: SPIE Society of Photo-Optical Instrumentation Engineers

 Fecha de publicación: 28/05/2013

Nº de páginas: 12

Tipo de publicación: Comunicación a Congreso

 DOI: 10.1117/12.2017458

ISSN: 0277-786X,1996-756X