Abstract: Real-time tissue identification during laser-guided surgery is a critical unmet need for collateral damage voidance and margin delineation. Laser-Induced Breakdown Spectroscopy (LIBS) is compatible with pulsed laser surgical systems and offers rapid, label-free elemental analysis. This study presents a machine learning pipeline classifying five ex vivo bovine tissue classes, plus one synthetic null-signal control class, from LIBS spectra, designed to control specimen-level data leakage and class imbalance bias. Key contributions
are (i) a ´peak max over baseline? aggregation strategy suppressing shot noise while preserving emission peaks; (ii) a repeated, group-based cross-validation protocol (GroupShuffleSplit, N = 10) enforcing specimen-level separation; and (iii) a comparison of 30 configurations (10 classifiers × 3 pipelines). Extra Trees with normalization reached the highest weighted F1-score (0.934 ± 0.118); excluding the synthetic control, five-class scores fall to 0.875-0.915 and the ranking changes, so these are the reference figures for biological
tissue discrimination. Support Vector Machines were less accurate but more consistent (0.917 ± 0.069). Acquisition takes approximately 3 s per point; inference is sub-millisecond. With five source animals, the best configuration chosen on the same outer splits, and inner tuning that was not group-aware, these estimates are an exploratory step toward real-time guided surgery.
Autoría: Sosa-Santos R.F., Arce-Diego J.L., Fanjul-Vélez F.,
Fuente: Sensors, 2026, 26(16), 5020
Editorial: MDPI
Fecha de publicación: 07/08/2026
Nº de páginas: 24
Tipo de publicación: Artículo de Revista
DOI: 10.3390/s26165020
ISSN: 1424-8220
Proyecto español: PID2021-127691OB-I00