Temperature effects of imperfectly formed metal-ingots in high temperature fixed point crucibles

Abstract: This paper investigates, through thermal modeling the effect on the radiance temperature of imperfectly filled thermometric fixed points. A two dimensional axisymmetric thermal model was established and the effect on the radiance temperature of various ingot imperfections such as voids and cracks in different places and of different dimensions in the ingot structure was calculated. It was found that the radiance temperature of the fixed point is quite tolerant to even relatively large flaws in the ingot structure. Only when flaws of significant dimensions near the radiating back wall were introduced was the overall radiance temperature significantly affected.

Otras publicaciones de la misma revista o congreso con autores/as de la Universidad de Cantabria

 Fuente: Measurement: Sensors, 44(4), 738-742

Editorial: Elsevier

 Fecha de publicación: 01/05/2011

Tipo de publicación: Artículo de Revista

 DOI: 10.1016/j.measurement.2011.01.003

ISSN: 2665-9174

Url de la publicación: https://doi.org/10.1016/j.measurement.2011.01.003

Autoría

LEVICK, A.

VILLAMAÑÁN, MIGUEL ÁNGEL