Estamos realizando la búsqueda. Por favor, espere...
1549
37
172
26128
4138
2532
334
Abstract: The application of the compact disk (CD) method for radon measurements at mines, caves and other workplaces needs testing under highly variable exposure conditions. We present the results from a blind comparison of CDs exposed in the Laboratory of Natural Radiation (Saelices el Chico, Spain). During the exposure the temperature varied from 6.5 to 24.9 °C (average 12.6 °C) and the 222Rn activity concentrations varied from <10 Bq m-3 to 147 kBq m-3. Good correspondence was observed between the integrated 222Rn activity concentration determined by the reference instruments in the laboratory (122,500 ± 6100 kBq h m-3) and that assessed by analysis of the CDs at a depth 80 µm beneath the front surface (118,000 ± 12,000 kBq h m-3) and at a depth of 120 µm (106,000 ± 12,000 kBq h m-3). The theoretical modeling of the CD response under variable temperature and radon concentration suggested that the small bias is probably due to the time variation of the calibration factor because of the time variations of the temperature.
Fuente: Int J Environ Res Public Health, 16 (17) 2019 Aug 22
Editorial: MDPI
Año de publicación: 2019
Nº de páginas: 9
Tipo de publicación: Artículo de Revista
DOI: 10.3390/ijerph16173038
ISSN: 1661-7827,1660-4601
Url de la publicación: https://www.doi.org/10.3390/ijerph16173038
Consultar en UCrea Leer publicación
PRESSYANOV, DOBROMIR
LUIS SANTIAGO QUINDOS PONCELA
GEORGIEV, STRAHIL
DIMITROVA, IVELINA
MITEV, KRASIMIR
CARLOS SAINZ FERNANDEZ
ISMAEL FUENTE MERINO
DANIEL RABAGO GOMEZ
Volver